Missed Meal Bolus Reminders on the t:slim X2 Insulin Pump

If you have enabled Missed Meal Bolus reminders on your t:slim X2 insulin pump, the pump will notify you if a Food or Correction Bolus was not delivered during your programmed time period(s).

There are four separate time periods available for programming the Missed Meal Bolus reminder. Use the following steps to program or edit a reminder:

  1. From the Home screen, tap OPTIONS and then My Pump
  2. Tap Alerts & Reminders and then Pump Reminders
  3. Tap Missed Meal Bolus
  4. Tap Press to Turn On for one of the four time periods
  5. Choose the Selected Days, Start Time, and End Time
  6. Tap the green checkmark to save the details
  7. Repeat steps 4-6 for the remaining time periods or return to the Home screen

My Pump and Reminder 1 Screens. Note: Not all screens are shown.

Unwanted Reminders

If you enter an amount of units into the top center box of the bolus screen, instead of entering grams under the CARBS category and/or a blood glucose value under the GLUCOSE category, the t:slim X2 insulin pump will register the bolus as an Override Bolus, rather than a Food or Correction Bolus. As a result, the Missed Meal Bolus reminder could still occur. To prevent this in the future, be sure to also enter your carb amount on the Bolus screen.

Back to All Articles

Not finding the answers you need?

Contact Support