To display your current glucose reading and activate Control-IQ technology on your Tandem Diabetes Care insulin pump, you will need to connect your pump with a compatible CGM sensor.*
Tandem Diabetes Care and Dexcom have created life-changing diabetes management solutions with more than a decade of collaboration. Here is a breakdown of which Dexcom CGM sensors are currently compatible with your Tandem insulin pump.
Dexcom G7
Tandem pumps (software version 7.7 or later) are compatible with specific versions of the Dexcom G7 sensor.
Dexcom G6
Tandem pumps (software version 7.3 or later) are compatible with all versions of the Dexcom G6 sensor.
If the side of the box has a white line below the LBL number (see image) or a manufacturing date of 2024, then the Dexcom G7 sensor is compatible with your pump.
/ ! \ Important
If you receive a "Sensor Not Started (43T)" alert, you are attempting to pair with an incompatible sensor.
Please note: The numbers on your Dexcom G7 sensor box will be different than the numbers shown in the reference example on this page.
If you don’t have the box that your Dexcom G7 sensor came in, but your sensor is connected to the Dexcom G7 app on your smartphone, please use the following instructions to check compatibility.
Open your Dexcom G7 app and tap Connections.
Tap Sensor to continue.
Confirm that your Firmware version is 32.192.105.64 or later.