If you need to stop using your Tandem Mobi insulin pump for a period of time, you'll need to place the pump in storage mode. Please use the following steps:
- Place the Tandem Mobi pump on its charging pad
- Press and hold down the Pump button for approximately 20 seconds. The pump will beep three times before going into storage mode.
- Remove the pump from the charging pad
- Keep the pump protected when it is not in use
To turn your Tandem Mobi insulin pump back on, simply place the pump back on its charging pad and hold down the Pump button for approximately five seconds. The pump will beep four times and the status lights will flash green four times. If the pump has been in storage mode for an extended period, it may take longer than usual to turn back on.
Note: All of your Personal Profile information will be saved while your Tandem Mobi insulin pump is turned off, but your Insulin on Board (IOB) will not. When your pump is turned back on, you will need to load a new cartridge to resume insulin delivery.